读取测量值: 发送: xx 04 00 00 00 03 crcL crcH 应答: xx 04 06 v3 v2 v1 v0 tt rr crcL crcH 假设读值 V = 100000, 当 rr = 0xA4 时,即 200mR 档,分辨率 0.001 mR, R = V * 0.001 = 100.000 mR rr = 0xA5 时,即 2R 档,分辨率 0.01 mR,则 R = 1000.00 mR = 1.00000 R rr = 0xA6 时,即 20R 档,分辨率 0.1 mR,则 R = 10000.0 mR = 10.0000 R rr = 0xAC 时,即 200R 档,分辨率 0.001 R,则 R = 100.000 R rr = 0xAB 时,即 2KR 档,分辨率 0.01 R,则 R = 1000.00 R = 1.00000 KR rr = 0xAA 时,即 20KR 档,分辨率 0.1 R,则 R = 10000.0 R = 10.0000 KR rr = 0xA9 时,即 200KR 档,分辨率 1 R,则 R = 100000 R = 100.000 KR rr = 0xA8 时,即 2MR 档,分辨率 0.01 KR,则 R = 1000.00 KR = 1.00000 MR rr = 0xA7 时,即 20MR 档,分辨率 0.1 KR,则 R = 10000.0 KR = 10.0000 MR rr = 0x98 时,即 200MR 档,分辨率 1KR,则 R = 100000 KR = 100.000 MR rr = 0x99 时,即 2GR 档,分辨率 0.01 MR rr = 0x9A 时,即 20GR 档,分辨率 0.1 MR rr = 0x9B 时,即 200GR 档,分辨率 1 MR rr = 0x9C 时,即 2TR 档, 分辨率 0.01 GR rr = 0x9D 时,即 20TR 档,分辨率 0.1 GR 设置量程: 发送: xx 06 00 45 00 rr crcL crcH 应答: xx 06 00 45 00 rr crcL crcH rr 取值范围 rr = 0, 测量关闭,切回内部自检 rr = 0x20, 高阻自动量程 rr = 0x21, 2MR 量程 rr = 0x22, 20MR rr = 0x23 200MR rr = 0x24 2GR / 2000MR rr = 0x25 20GR rr = 0x26 200GR rr = 0x27 2TR / 2000GR rr = 0x28 20TR rr = 0x10, 低阻自动量程 rr = 0x11, 40mR 量程 rr = 0x12, 200mR rr = 0x13 2R rr = 0x14 20R rr = 0x15 200R rr = 0x16 2KR rr = 0x17 20KR rr = 0x18 200KR rr = 0x19 2MR 设置电压: 发送: xx 06 00 46 vH vL crcL crcH 应答: xx 06 00 46 vH vL crcL crcH 输出电压 Volt = vH * 256 + vL, 取值范围 0 ~ 1050 V 设置电流百分比(仅适用于最小的三个低阻档): 发送: xx 06 00 47 vH vL crcL crcH 应答: xx 06 00 47 vH vL crcL crcH 百分比 Percent = vH * 256 + vL, 取值范围 0 ~ 1000,表示 0.0% ~ 100.0%